Viscotek RImax - specifications
The Viscotek RImax is a complete, advanced, multidetector GPC/SEC system suitable for all macromolecular applications.
OmniSEC
- Acquisition
- 6 channels, 5Hz, unlimited acquisition time
- Calculation modules
- Conventional, Universal, Light Scattering, Triple, Copolymer, Protein specific, FIPA
- Results output
- Mn, Mw, Mz, Mp, Mw/Mn, IV [η], Rh, Rg, Branching Number, Branching Frequency, Number of Arms, Weight Fraction, Concentration, MH-a, MH-k, dn/dc, dA/dc. A2
- Minimum PC specification
- Windows XP or higher, 100 GB HD, 2GB RAM, 1.8 GHz processor,
Graphics 1024x768, 16-bit colour depth
Refractive index detector
- Measuring principle
- Deflection
- Light Source
- Tungsten lamp
- Cell
- 45° quartz glass, 9μL volume
- Purging
- 3 modes, Manual, Automatic and Programmable
- Data rate
- 5Hz to OmniSEC via 270 (2Hz direct to OmniSEC in stand-alone mode)
- Temperature control
- Normal environment, 35 to 55ºC in 1ºC steps.
Cold environment, 10 to 30ºC in 1ºC steps. - Dimensions
- 22cm x 16cm x 35cm (W x H xD); 12kg
Column Oven
- Temperature control
- 10ºC above ambient to 100ºC
- Capacity
- Up to five 8mm x 30cm GPC columns
- Internal dimensions
- 41cm x 7cm x 4cm (W x H xD)
UV detector
- Wavelength
- 190 to 740nm
- Light Source
- Deuterium; Tungsten halogen is optional
- Cell volume
- 10μL
- Data rate
- Analog collection of single wavelength (programmable), A-to-D with 5Hz to OmniSEC
- Dimensions
- 23cm x 14cm x 39cm (W x H xD); 6.0kg
UV-PDA detector
- Wavelength/resolution
- 190 to 500nm or 430 to 710nm on 256 diode array
- Light Source
- Deuterium (190-500nm) or Tungsten (430-710nm)
- Cell volume
- 10μL
- Data rate
- 256 channels at 1Hz direct to OmniSEC
- Dimensions
- 23cm x 14cm x 39cm (W x H xD); 6.0kg
GPCmax
- Operation and Control
- By OmniSEC software or from front panel
- Dimensions
- 55cm x 38cm x 53cm (W x H x D); 37kg
Degasser
- Channels/volume
- Two channels, 8ml per channel
- Performance
- <0.5 ppm oxygen at 0.5 mL/min
- Eluent Sensor
- User selectable, stops GPC pump when eluent runs out
Pump
- Flow rate range
- 0.01 to 9.99 mL/min
- Pulsation
- Less than 1% (measured with viscometer)
- Pressure Reading
- MPa or PSI
- Soft start/stop
- User programmable in mL/min/min
Autosampler
- Capacity (Standard)
- 120 vials, freely programmable position sequence
- Range
- Variable injection volume, between 20 and 150μL
- Reproducibility
- Better than 0.5 %
- Carryover
- Less than 0.1%, depending on wash program
- Cooling option
- 4°C to 40°C
- Capacity (with cooling option)
- 60 vials cooled, freely programmable position sequence
- Heated tray option
- 65°C fixed in one tray
- Capacity (with heating option)
- 60 vials heated, 60 unheated, freely programmable position sequence